GENERAL FEATURES
LCD display screen, 3.5 inches.
Multi-reading feature allows auto-read and continuous-read.
Automatic/Manual temperature compensation ensures accurate results.
Auto-hold feature senses and locks the measurement endpoint.
Data capacity of up to 200 sets for each parameter.
Auto-power off feature effectively extends the battery service life.
Reset feature automatically resumes all settings back to factory default options.
IP65 waterproof.
1 point calibration with Standard Recognition.
Cell constant is settable
Automatic/manual temperature compensation ensures accurate results.
PRODUCT SPECIFICATIONS
Model | EC210E | |
Parameters | EC/Temp.(TDS) | |
Conductivity | Range | 0.00 μS/cm to 200 mS/cm |
Resolution | 0.01 μS/cm minimum, various with range selection | |
Accuracy | ±1.0% FS | |
Reference Temperature | 25 ºC | |
Calibration Points | Up to 1 | |
Standard Recognition | 84μS/cm, 1413μS/cm, 12.88mS/cm | |
TDS | Range | 0.00mg/L~300g/L |
Resolution | 0.01mg/L minimum, various with range selection | |
Accuracy | ±1.0%FS | |
Temperature | Range | -5 to 110 ºC, 23 to 230 ºF |
Unit | ºC, ºF | |
Resolution | 0.1 | |
Relative Accuracy | ±0.2 | |
Measurement | Reading Mode | Auto Read, Continuous |
Reading Prompts | Reading, Stable, Locked | |
Temp. Compensation | ATC, MTC | |
Data management | Data Storage | 200 results each |
Inputs | Conductivity with Temp. Probe | 5-pin aviation connector |
Display options | Backlight | Yes |
Auto Shutdown | 300, 600, 1200, 1800, 3600 sec., off | |
IP Rating | IP65 | |
General | Power | Rechargeable Lithium battery, AC Adapter, 100-240V AC input, DC5V output |
Dimensions | 80×255×35 mm | |
Weight | 400g (0.88 lb) |