Home > News Center > Technical articles

Using a pH meter with ease—a must-read

The pH meter is mainly used to measure the acidity and alkalinity levels in water bodies. It determines the acidity of water through electrochemical or chemical indicator methods and is used for water quality monitoring and environmental protection.

 handheld ph meter

Understanding these points will make using lab pH meter much easier:

1. Why are the pH values of the same sample different in two measurements?

Temperature changes or chemical reactions within the sample itself can cause pH changes. Therefore, it's important to maintain a consistent temperature and avoid chemical reactions.

2. Why are the readings inconsistent when measuring the same sample simultaneously on two different pH meters?

This is because the calibration conditions of the two pH meters are different, resulting in different measured values. Therefore, the pH meters should be calibrated using the same buffer solution at the same time before simultaneous measurements.

3. Why has the buffer solution deteriorated and become unusable within its expiration date?

The expiration date of a buffer solution refers to its shelf life in its unopened state. Once opened, the buffer solution is more prone to deterioration.

4. How often should electrodes be calibrated?

The calibration frequency of electrodes depends on specific factors such as electrode usage, maintenance, sample properties, and measurement accuracy. It is recommended to calibrate daily, and no more than once a week.

 conductivity ph meter

NANBEI INSTRUMENT has earned the trust of universities, research institutions and numerous enterprises, with services spanning fields of chemical, pharmaceuticals, testing, agriculture, food, metallurgical mining, construction and petroleum.