Home > Water Analysis Equipments > Multi–Parameter Meter > Bench Type Multi-Parameter

M500T Multi-parameter Analyzer

M500T Multi-parameter Analyzer

M500T Multi-parameter Analyzer

M500T Multi-parameter AnalyzerGENERAL FEATURES:Color High contrast LCD touchscreen, 7.0 inches.Intelligent operation sys
Contact Now
Detailed Product Description

M500T Multi-parameter Analyzer

GENERAL FEATURES:

Color High contrast LCD touchscreen, 7.0 inches.

Intelligent operation system provides features including user management, calibration management, method management, data management and etc.

Multi-reading feature allows auto-read, timed-read and continuous-read.

Automatic/Manual temperature compensation ensures accurate results.

Data Storage 1000 sets for each parameter (GLP-compliant).

Data analysis feature helps user review, compare and recalculate results.

Support for USB or RS-232 communication.

Reset feature automatically resumes all settings back to factory default options.

IP54 waterproof.

 

pH

1-6 points calibration with Standard Recognition.

Selectable pH buffer groups, including DIN, NIST, USA, MERK, JIS, GB.

Support for pH standard group customization.

Automatic electrode diagnosis with pH slope and offset display.

 

Ion

1-6 points calibration

Selectable measurement unit, including μg/L, mg/L, g/L, mmol/L, pX, ppm, etc.

Multi-measurement modes are supported, including Direct Reading mode, Standard Addition mode, Sample Addition mode and GRAN mode.

Over 10 methods are built-in, including F-, Cl-, Br-, I-, NO3-, BF4-, NH4+, K+, Na+, Ca2+, Cu2+, Pb2+, Ag+ and etc., user-defined method is supported.

 

Conductivity

1-5 points calibration with Standard Recognition.

Settable parameters, including cell constant, reference temperature (5/10/15/18/20/25°C), temperature compensation coefficient and TDS factor.

Temperature compensation type (none, linear, pure water).

 

DO

Support for air-saturated water or zero oxygen calibration.

Auto barometric pressure compensation.

Manual Salinity Factor Correction.

Selectable pressure unit, including kPa, mbar, Torr, Atm.

 

PRODUCT SPECIFICATIONS

Model

M500T

Parameters

pH/EC/ISE/DO/Temp. (mV/ORP/pX/Resistivity/TDS/Sal./DO Saturation)

pH

Range

-2.000 to 20.000 pH

Resolution

0.1, 0.01, 0.001pH

Accuracy

±0.002 pH

Calibration Points

Up to 6

Standard Customization

Yes

Calibration Reminder

Yes

Standard Recognition

NIST, GB, DIN, USA, MERK, JIS buffers

Slope Limit

Yes

mV/ORP

Range

-2000.00 to 2000.00 mV

Resolution

0.1, 0.01 mV

Accuracy

±0.1 mV or ±0.03% of

reading whichever is greater

EH ORP Mode

Yes

ORP Calibration Points

1 custom point (relative mV)

pX

Range

-2.000 to 20.000 pX

Resolution

0.1, 0.01, 0.001 pX

Accuracy

±0.002 pX

Calibration Points

Up to 6

ISE

Range

1E-9 to 9.999E9

Units

mol/L, mmol/L, g/L, mg/L, μg/L, ppm

Resolution

Up to 4 significant digits

Accuracy

±0.3%

Calibration Points

Up to 6

Conductivity

Range

0.000 μS/cm to 2000 mS/cm

Resolution

0.001 μS/cm minimum,

various with range selection

Accuracy

±0.5% FS

Reference Temperature

5, 10, 15, 18, 20, 25 ºC

Calibration Points

Up to 5

Calibration Reminder

Yes

Standard Recognition

10 μS/cm, 84 μS/cm, 500 μS/cm, 1413 μS/cm, 12.88 mS/cm

Resistivity

Range

5.00 Ω·cm~100.00 MΩ·cm

Resolution

0.01 Ω·cm minimum

Accuracy

±0.5% FS

TDS

Range

0.000~1000 ppt

Resolution

0.001 ppm minimum, changed with range

Accuracy

±0.5%FS

Salinity

Type

NaCl%

Range

(0.00~8.00)%

Resolution

0.01%

Accuracy

±0.1%

Dissolved

Oxygen Concentration - Polarographic

Range

0.00 to 99.99 mg/L

Resolution

0.01 mg/L

 

Accuracy

±0.3mg/L (0.00 to 20.00 mg/L);

±10% (20.00 to 99.99 mg/L)

Calibration Reminder

Yes

Calibration Points

Air-saturated water or zero point

Barometric Compensation

Yes

Manual Salinity Factor Correction

Yes

% Saturation -Polarographic

Range

 (0.0 to 600)%

Resolution

0.10%

 

Accuracy

±2.0%

Temperature

Range

-10 to 135 ºC, 14 to 275 ºF

Unit

ºC, ºF

Resolution

0.1

Accuracy

±0.1

Measurement

Reading Mode

Auto Read (Fast, Medium, Slow), Timed, Continuous

Reading Prompts

Reading, Stable, Locked

Temp. Compensation

ATC, MTC

Data management

Data Storage

1000 Groups

GLP Features

Yes

Log Management

Yes

Inputs

pH Electrode

BNC(Q9)

DO with Temp. Probe

4-pin aviation connector

Conductivity with Temp. Probe

5-pin aviation connector

Outputs

USB

USB 2.0 flash memory device, PC

RS 232

printer, scanner, auto-sampler

Display options

Backlight

Yes

Auto Shutdown

1~60 min, off

IP Rating

IP54

Date and Time

Yes

General

Power

AC Adapter, 100-240V AC input, DC24V output

Dimensions

220×195×68 mm

Weight

950g (2.09 lb)


Contact Details
Other Products